1 general information on the msm‑hc, 2 connecting the msm‑hc, 18 high capacity metrohm suppressor module – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual
Page 72: Msm-hc), General information on the msm-hc, Connecting the msm-hc, See chapter 3.18
3.18 High Capacity Metrohm Suppressor Module (MSM-HC)
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
850 Professional IC Anion – MSM-HC – MCS – Prep 3
3.18
High Capacity Metrohm Suppressor Module
(MSM-HC)
3.18.1
General information on the MSM-HC
The MSM-HC is used for chemical suppression during anion analysis if elu-
ents with a high ionic concentration are used or if the concentration is
being significantly increased in the course of a gradient application. It con-
sists of 3 suppressor units in total, which are used for suppression, regen-
erated with sulfuric acid or rinsed with ultra pure water in rotation.
Suppression reaction in the MSM-HC
When using a carbonate eluent, the following reaction (amongst others)
occurs in the MSM-HC:
R-SO
3
-
H
+
+ NaHCO
3
/Na
2
CO
3
➙ R-SO
3
-
Na
+
+ H
2
O + CO
2
3.18.2
Connecting the MSM-HC
Caution
To protect the MSM-HC against foreign particles or bacterial growth, a
6.2744.180 pump tubing connection with filter must be mounted
between the peristaltic pump and the inlet capillaries of the MSM-HC.
The three inputs and outputs of the suppressor units numbered with 1, 2
and 3 on the MSM-HC each have 2 fixed mounted PTFE capillaries (see
Figure 37, page 63).