beautypg.com

16 mains connection – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual

Page 136

background image

7.16 Mains connection

■■■■■■■■■■■■■■■■■■■■■■

126

■■■■■■■■

850 Professional IC Anion – MSM-HC – MCS – Prep 3

Conductivity
detector

Cell volume

0.8 µL

Cell constant

Individual calibration data saved in the detector

adjustable range: 13.0…21.0 /cm

Electrodes

Ring-shaped electrodes made from stainless steel

Materials in
contact with
eluent

Chemically inert PCTFE

Maximum oper-
ating pressure

5.0 MPa (50 bar)

Cell tempera-
ture

20…50 °C in increments of 5 °C

Temperature
stability

< 0.001 °C

Temperature
compensation

0…5 %/K adjustable, default 2.3 %/K

Heating up time

< 30 minutes (40 °C)

7.16

Mains connection

Required voltage

100…240 V ± 10 % (auto-sensing)

Required fre-
quency

50…60 Hz ± 3 (auto-sensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

internal fuse 3.15 A