5.14 Metrohm CO

2

suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

112

■■■■■■■■

850 Professional IC Anion – MSM-HC – MCS – Prep 3

5.14

Metrohm CO

2

suppressor (MCS)

5.14.1

Replacing the CO

2

adsorption cartridge

The CO

2

adsorption cartridge 6.2837.000 (39-4) must be replaced regu-

larly. This is because of blockages and losses in capacity.

Blockage
Moisture can block the CO

2

adsorption cartridge. This is revealed by a

change in color of the cartridge material (the orange part becomes color-
less). As the air flow is reduced, the vacuum becomes lower. To protect
the CO

2

adsorption cartridge, an H

2

O adsorption cartridge (39-7) is instal-

led upstream. Regular regeneration (see Chapter 5.14.2, page 112) of
the H

2

O adsorption cartridge extends the service life of the CO

2

adsorp-

tion cartridge.

Capacity loss
The adsorption capacity of the CO

2

adsorption cartridge is limited.

Depending on the period of operation and laboratory environment, the
adsorption capacity will diminish over time. This is manifested in a rising
baseline (as more CO

2

reaches the detector).

5.14.2

Regenerating the H

2

O adsorption cartridge

The function of the H

2

O adsorption cartridge is to protect the CO

2

adsorp-

tion cartridge against moisture. The service life of the H

2

O adsorption car-

tridge depends on the moisture content of the ambient air. Moisture
reduces the capacity of the H

2

O adsorption cartridge (which can be

observed by a change in color). Before the color changes in the entire fill-
ing material (from orange to colorless, with Fluka Order No. 94098), the
H

2

O adsorption cartridge should be regenerated (see information sheet).

The filling material is replaced during regeneration:

1 Allow material to dry loose (not in cartridge) at 140 °C overnight and

refill. Or dispose of the old material, and fill with new material.

2 Cover the packed material with wadding.