3.14 Sample preparation module (SPM)

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

850 Professional IC Anion – MSM-HC – MCS – Prep 3

Weak alkaline solutions can directly be forwarded to the SPM as with cat-
ion exchange. However, strong alkaline solutions (e.g. 30 % NaOH) could
exceed the capacity of the SPM. In order to ensure a sufficient capacity for
exchanging the Na

+

ions for H

+

ions in all cases, only a small part of the

sample (e.g. 20 µL) is therefore given onto the SPM. For measuring the
sample the sample loop of an additional injection valve (27-A) (e.g. on
the Sample Processor) is needed.

Procedure of the neutralization with the SPM

1 Measuring the sample

Valve positions:

Injection valve 1 (27-A) in position Fill.
– Injection valve 2 (in the instrument) in position Inject.

On the injection valve 1 (27-A) the sample is carried through the
sample loop and is measured.

1

A

B

Figure 27

Measuring the sample

1

Sample

A

Injection valve 1
With sample loop. For measuring the sam-
ple.

B

Injection valve 2
With preconcentration column. For precon-
centrating and injecting the analyte ions.

2 Neutralization and preconcentration

Switch both injection valves:

Injection valve 1 (28-A) to position Inject.
Injection valve 2 (28-C) to position Fill.

The sample is carried with the transfer solution (28-2) onto the
SPM (28-B).
Na

+

ions are exchanged for H

+

ions.

The analyte ions and the transfer solution flow to the injection
valve 2.