6 Troubleshooting

850 Professional IC Anion – MSM-HC – MCS – Prep 3

■■■■■■■■

119

Problem

Cause

Remedy

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 75)
.

Conductivity detec-
tor is not recognized
in the software

No connection.

Check the cable connection (41-

1

).

Turn the instrument off and on again (after
15 seconds).

Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)

Connections – dead vol-
ume in the system.

Check the connections (see Chapter 3.5, page
19)
(use PEEK capillaries with an internal diam-
eter of 0.25 mm between the injection valve
and detector).

Guard column – diminished
performance.

Replace guard column (see Chapter 3.22,
page 72)
.

Separation column – dead
volume at column head.

Install separation column in reverse flow
direction (if permitted by the information
sheet).

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 75)
.

Greater rise in the
baseline

SPM – reduced capacity.

Regenerate the SPM (see Chapter 5.10.3.1,
page 98)
.

MSM-HC – reduced
capacity.

Regenerate the MSM-HC (see Chapter
5.13.3.1, page 107)
.

Precision problems -
significant scatter-
ing of the measured
values

Sample – Gas bubbles in
the sample.

Use sample degasser (see Chapter 3.13, page
45)
.

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 3.16.1, page 57)
.

Sample – Rinsing volume
too low.

Prolong the rinsing time (see Chapter 5.8,
page 95)
.

Injection valve – defective.

Request Metrohm Service.

MCS – vacuum too low.

Check connectors. If they are ok:

Contact the Metrohm Service.