Figure 25, Cation exchange with the spm, Figure 26 – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual

Page 59: Injection and sequential suppression

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC Anion – MSM-HC – MCS – Prep 3

■■■■■■■■

49

1

A

B

D

C

Figure 25

Cation exchange with the SPM

1

Sample

A

SPM

B

Injection valve

C

Separation column

D

MSM-HC

2 Injection and sequential suppression

The injection valve is switched to the position Inject.

With the eluent flow (26-2) the sample prepared is rinsed onto
the separation column (26-C)
.

The sequential suppression with the MSM-HC (26-D) and the
MCS follows after the separation.

A

B

D

C

2

Figure 26

Injection and sequential suppression

2

Sample in the eluent flow

A

SPM

B

Injection valve

C

Separation column

D

MSM-HC

Neutralization
A typical use of the SPM is the neutralization of an alkaline sample for
anion determination with chemical suppression.