3.18 High Capacity Metrohm Suppressor Module (MSM-HC)
■■■■■■■■■■■■■■■■■■■■■■
64
■■■■■■■■
850 Professional IC Anion – MSM-HC – MCS – Prep 3
Connecting the capillaries of the MSM-HC
1 Connecting the eluent inlet capillary
■
Fasten the end of the inlet capillary labeled with Eluent with a
6.2744.070 short PEEK pressure screw to the output of the col-
umn.
2 Connecting the eluent outlet capillary
■
Fasten the end of the outlet capillary labeled with Detector with a
long PEEK pressure screw (6.2744.090) to the input of the MCS.
OR
Connect the end of the outlet capillary labeled with Detector and
the detector inlet capillary with a 6.2744.040 coupling and two
6.2744.070 short pressure screws.
3 Connecting the rinsing solution inlet capillary
■
Fasten the end of the inlet capillary labeled with H20 with a
6.2744.070 short PEEK pressure screw to the pump tubing con-
nection of the pump tubing which carries the rinsing solution.
4 Connecting the rinsing the solution outlet capillary
■
Guide the other end of the outlet capillary labeled with Waste
into a sufficiently large waste container and fasten it there.
5 Connecting the regeneration solution inlet capillary
■
Fasten the end of the inlet capillary labeled with H2S04 with a
6.2744.070 short PEEK pressure screw to the pump tubing con-
nection of the pump tubing which carries the regeneration solu-
tion.
6 Connecting the regeneration solution outlet capillary
■
Guide the other end of the outlet capillary labeled with Waste
into a sufficiently large waste container and fasten it there.
The rinsing and regeneration solutions are pumped with a peristaltic pump
(see Chapter 3.15, page 52).