beautypg.com

3 replacing parts of the spm – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual

Page 111

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

850 Professional IC Anion – MSM-HC – MCS – Prep 3

■■■■■■■■

101

7 Inserting the SPM connecting piece

Insert the SPM connecting piece (53-2) into the SPM housing
(53-4) in such a way that the connector 1 is on top and the three
pins of the SPM connecting piece fit into the corresponding
recesses on the SPM housing (53-4).

8 Connecting and conditioning the SPM

Reconnect the SPM to the IC system.

Before switching the SPM over for the first time, rinse the three
neutralization units with solution for 5 minutes.

5.10.3.3

Replacing parts of the SPM

It may be necessary to replace parts of the SPM in the following cases:

Loss of neutralization capacity which cannot be eliminated (reduced
detection sensitivity).

Blockage of the SPM which cannot be eliminated (solutions can no lon-
ger be pumped through the SPM).

Both the SPM rotor A (53-3) and the SPM connecting piece (53-2) with its
supply and discharge lines can be replaced.

Replacing parts of the SPM

Replace parts of the SPM as follows (see Figure 53, page 99):

1 Disconnecting the SPM from the IC system

Switch off the instrument.

Disconnect capillary connections of the SPM from the system.

2 Dismounting SPM

Unscrew union nut (53-1) from the SPM housing (53-4).

Pull the SPM connecting piece (53-2) and the SPM rotor A (53-3)
out of the SPM housing (53-4). The SPM connecting piece and
the SPM rotor A normally stick to one another - if this is not the
case: Take a sharp object, insert into the slot (53-5) in the SPM
housing, and pull out the SPM rotor A (53-3) in this way.

Detach the SPM connecting piece (53-2) from the SPM rotor A
(53-3).

3 Cleaning the new SPM rotor A

Clean sealing surface of the new SPM rotor A (53-3) with ethanol
using a lint-free cloth.