9.2 Reagent monitoring
■■■■■■■■■■■■■■■■■■■■■■
68
■■■■■■■■
915 KF Ti-Touch
■
at the end of the determination.
Number of determ.
The number of determinations to be carried out with a certain amount of
reagent depends on the type of sample and its amount.
Input range
1 - 999
Selection
off
Default value
off
Working life
Working life of the reagent.
Input range
1 - 999 days
Selection
off
Default value
off
Volume
Volume of titrant dosed.
Input range
1.0 - 999.9 mL
Selection
off
Default value
off
[Status]
Display the status overview of the current values of the reagent monitor-
ing.
[Reagent replacement]
Edit the parameters for the reagent exchange.
Dialog "Reagent monitoring / Status"
The current reagent monitoring values are displayed in this dialog.
[Reset]
Reset the values to zero.
Dialog "Reagent monitoring / Reagent replacement"
The parameters for the reagent exchange are defined in this dialog.
Reagent replacement
The reagent can either be exchanged manually or automatically.
Selection
manual | auto
Default value
manual
manual
If a monitored parameter has reached the limit set, a message is being
displayed. Then the reagent has to be exchanged manually.