2 control parameters, Control parameters – Metrohm 915 KF Ti-Touch User Manual
Page 275

■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
915 KF Ti-Touch
■■■■■■■■
261
28.1.2
Control parameters
The control parameters for the endpoint are defined under [Control
parameters].
Endpoint at
Measured value for the endpoint.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 355.
Measuring mode Ipol:
Input range
–2,000.0 - 2,000.0 mV
Default value
250.0 mV
Selection
off
Measuring mode Upol:
Input range
–200.00 - 200.00 µA
Default value
25.00 µA
Selection
off
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection
slow | optimal | fast | user
Default value
optimal
slow
For samples with a low water content or samples which release their
moisture only slowly.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For uncritical samples with high water content.
user
The individual titration parameters can be modified.
The settings of the individual titration rates are listed in table 6, page
263.
User-defined parameters
These parameters are only accessible when Titration rate = user.
Dynamics
This parameter defines the control range before the specified endpoint.
Individual volume steps are dosed in the control range, the dosing is finely