28 Parameters
915 KF Ti-Touch
■■■■■■■■
259
28 Parameters
NOTE
For most of the numerical parameters, the result of a calculation can
also be utilized in place of a number. A description of this can be found
in the appendix (see Chapter 31.6, page 344).
28.1
Volumetric Karl Fischer titrations (KFT)
28.1.1
Start conditions
The parameters that are carried out before the start of titration are
defined under [Start conditions].
Pause 1
Waiting time, e.g. for the electrode to settle down before a start volume is
added.
Input range
0 - 999999 s
Default value
0 s
Start volume
Volume that is dosed prior to the start of the titration.
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Dosing rate
Rate at which the start volume is dosed.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 338).
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.
Input range
0 - 999999 s
Default value
0 s