28 parameters, 1 volumetric karl fischer titrations (kft), 1 start conditions – Metrohm 915 KF Ti-Touch User Manual

Page 273: 1 volumetric karl fischer titrations (kft) 28.1.1, Start conditions

background image

■■■■■■■■■■■■■■■■■■■■■■

28 Parameters

915 KF Ti-Touch

■■■■■■■■

259

28 Parameters

NOTE

For most of the numerical parameters, the result of a calculation can
also be utilized in place of a number. A description of this can be found
in the appendix (see Chapter 31.6, page 344).

28.1

Volumetric Karl Fischer titrations (KFT)

28.1.1

Start conditions

The parameters that are carried out before the start of titration are
defined under [Start conditions].

Pause 1

Waiting time, e.g. for the electrode to settle down before a start volume is
added.

Input range

0 - 999999 s

Default value

0 s

Start volume

Volume that is dosed prior to the start of the titration.

Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 338)
.

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.

Input range

0 - 999999 s

Default value

0 s