27.3 Dosing
■■■■■■■■■■■■■■■■■■■■■■
244
■■■■■■■■
915 KF Ti-Touch
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 338).
Fill automatic.
on | off (Default value: on)
If this parameter is activated, then the dosing cylinder is being automati-
cally filled at the end of the dosing.
27.3.3
Preparing
Manual control: Dosing
▶ Prepare
The Prepare function is used to rinse and fill the cylinder and tubings of
the exchange or dosing unit air bubble-free. You should carry out this
function before the first determination or once per day.
If the titrant is selected (see "Selecting the dosing function", page 238),
the parameters defined for the titrant for preparing/emptying and the tub-
ing parameters will be used (dosing unit only). If the titrant is not selected,
default parameters will be used (see Chapter 31.1.2, page 338 and Chap-
ter 31.2.2, page 339).
The following two figures show the live display of an exchange unit and a
dosing unit:
Figure 12
Live display "Preparing the dosing unit"