28 Parameters
915 KF Ti-Touch
■■■■■■■■
307
Filling rate
Rate at which the dosing cylinder is filled.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 338).
28.6.3.2
Control device
The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System
▶ Device
manager.
NOTE
This button will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device.
Control device
Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command.
Selection
Selection of configured control devices
28.6.3.3
Dosing device
The parameters for the dosing device are edited under [Dosing device].
Dosing device
Selection of the MSB connector the dosing device is connected to. The
selection does not depend on the number of MSB connectors of the con-
trol device.
Selection
1 | 2 | 3 | 4
Default value
1
Titrant
Selection of the titrant from the titrant list. We recommend always select-
ing the titrant. This ensures that the correct data (titer, etc.) is always used
for the calculation. Titrants are defined under System
▶ Titrants. You
can also enter a name which is not contained in the titrant list. When a
determination is started there is a check whether the solution is contained
in the list.