3 titration parameters, Titration parameters – Metrohm 915 KF Ti-Touch User Manual
Page 278

28.1 Volumetric Karl Fischer titrations (KFT)
■■■■■■■■■■■■■■■■■■■■■■
264
■■■■■■■■
915 KF Ti-Touch
Stop drift
This parameter can only be edited with Stop criterion = drift.
The titration is stopped when the endpoint and the stop drift have been
reached.
Input range
1 - 999 µL/min
Default value
20 µL/min
Delay time
This parameter can only be edited with Stop criterion = time.
When the endpoint has been reached, the specified time is allowed to
elapse after the last dosing and the titration is then stopped.
Input range
0 - 999 s
Default value
10 s
Relative stop drift
This parameter can only be edited with Stop criterion = rel. drift.
The titration is stopped when the endpoint and the sum of the drift at the
start of the titration and the relative stop drift have been reached.
Input range
1 - 999 µL/min
Default value
10 µL/min
28.1.3
Titration parameters
Under [Titration parameters], the parameters influencing the sequence
of the entire titration are defined.
Titration direction
Selection
+ | - | auto
The default value depends on the measuring mode:
Ipol = -, Upol = +
+
Positive measured value modification, i.e. in the direction of a greater
voltage or greater current.
-
Negative measured value modification, i.e. in the direction of a lesser
voltage or lesser current.
auto
The titration direction is determined automatically from the start mea-
sured value and the set endpoint.