Metrohm 915 KF Ti-Touch User Manual
Page 350
![background image](/manuals/586302/350/background.png)
30.7 Karl Fischer titration
■■■■■■■■■■■■■■■■■■■■■■
336
■■■■■■■■
915 KF Ti-Touch
Problem
Cause
Remedy
A side reaction is taking
place.
■
Use special reagents.
■
Adjust the method (increase/decrease the
temperature, external extraction).
■
See technical literature.
The pH value is not in the
optimum range anymore.
Add buffer, see technical literature.
The titration will not
be finished.
The titration cell is leaking.
■
Check seals and septum. If necessary,
replace.
■
Replace the molecular sieve.
The minimum increment is
too low.
Define Titration rate = user and increase the
minimum volume increment (Min. incre-
ment) (see Chapter 28.1.2, page 261).
The stop criterion is unsuit-
able.
Adjust the control parameters (see Chapter
28.1.2, page 261):
■
Increase the stop drift.
■
Select a short delay time.
See also: The drift becomes
greater after each titration.
The sample is over-
titrated.
The increments at the end
of the titration are too
high.
■
Define Titration rate = user and reduce
the dosing rate (Max. rate) (see Chapter
28.1.2, page 261).
The following experiment provides a clue
for the optimum dosing rate: During condi-
tioning, display the drift and add sample
without starting the titration. Select a value
below the highest drift as dosing rate.
■
Stir faster.
The amount of methanol in
the working medium is too
low.
■
Replace the working medium.
■
Reduce the amount of solubility promoter,
if working with solvent mixtures, see tech-
nical literature.
The electrode may be cov-
ered.
Wipe off the electrode with ethanol or a suit-
able solvent.
The solution
becomes darker
after each titration.
Replace the working medium.
The electrode may be cov-
ered.
Wipe off the electrode with ethanol or a suit-
able solvent.