beautypg.com

Metrohm 848 Titrino plus User Manual

Page 98

background image

7.2 Monotonic equivalence point titrations (MET)

■■■■■■■■■■■■■■■■■■■■■■

88

■■■■■■■■

848 Titrino plus

not defined
No check takes place.

I(pol)

The polarization current is the current that is applied at a polarizable elec-
trode during the voltametric measurement. This parameter is available
only with I(pol) determinations.

Input range

–125 - 125 µA (Increment: 1)

Default value

5 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement. This parameter is available only
with U(pol) determinations.

Input range

–1250 - 1250 mV (Increment: 10)

Default value

400 mV

Electrode test

For polarizable electrodes, an electrode test can be carried out. A check is
made that the electrode is properly connected and that no short-circuit is
present. The electrode test is carried out when the determination is star-
ted. This parameter is available only with I(pol) and U(pol) determinations.

Selection

on | off

Default value

off

Stirrer

When this parameter is switched on, then the stirrer is switched on at the
start of the determination.

Selection

on | off

Default value

on

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 U/min. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 118. The optimum stirring
rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

–15 - 15

Default value

8