6 System settings

848 Titrino plus

■■■■■■■■

65

The interface parameters for the connected balance are set under COM1
settings
.

Baud rate

Transfer rate in characters per second.

Selection

1200 | 2400 | 4800 | 9600 | 19200 | 38400 |
57600 | 115200

Default value

9600

Data bits

Number of data bits.

Selection

7 | 8

Default value

8

Stop bits

Number of stop bits.

Selection

1 | 2

Default value

1

Parity

Type of parity testing.

Selection

even | none | odd

Default value

none

Handshake

Type of the data transfer protocol.

Selection

hardware | software | none

Default value

hardware

Note

If communications problems occur, set the parameter Handshake to
software, and make another attempt.

Editing the COM2 settings
Menu

System External devices COM2 settings

The interface parameters for instruments connected to the RS-232/2 con-
nector of the RS-232/USB Box (e.g. PC) are set under COM2 settings.
The parameters and input ranges are identical to those for the COM1
interface.