beautypg.com

2 editing a calculation – Metrohm 848 Titrino plus User Manual

Page 104

background image

7.2 Monotonic equivalence point titrations (MET)

■■■■■■■■■■■■■■■■■■■■■■

94

■■■■■■■■

848 Titrino plus

The result name is specified in the list for each calculation.

Edit

Edit the data of the selected calculation, see following chapter.

Delete

Delete the selected calculation.

7.2.5.2

Editing a calculation

Result name

The result name is the text which will be shown in the result view and in
the report.

Entry

12 characters

Default value

empty

R1=…R5=

Shows the calculation formula. A special editor is opened for the defini-
tion (see Chapter 5.3, page 23).

Entry

44 characters

Default value

empty

Decimal places

Number of decimal places used to display the result.

Input range

0 - 5

Default value

2

Result unit

The result unit is displayed and saved together with the result.

Selection

% | mol/L | mmol/L | g/L | mg/L | mg/mL | ppm |
g | mg | mL | mg/piece | °C | µL | mL/min | User-
defined

Default value

%

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.

Save as titer

The result can be saved as titer for this selected solution. If statistics has
been switched on, then the current mean value of the determination ser-
ies will be saved.

Selection

on | off

Default value

off