11 technical specifications, 1 measuring inputs, 1 potentiometry – Metrohm 848 Titrino plus User Manual
Page 140: 2 polarizer, 1 measuring inputs 11.1.1, Potentiometry, Polarizer
11.1 Measuring inputs
■■■■■■■■■■■■■■■■■■■■■■
130
■■■■■■■■
848 Titrino plus
11 Technical specifications
11.1
Measuring inputs
The measuring cycle is 100 ms for all measuring modes.
11.1.1
Potentiometry
A high-ohm measuring input (Ind.) for pH and redox electrodes and a
measuring input for separate reference electrodes (Ref.).
Input resistance
> 1 · 10
12
Ω
Offset current
< 1 · 10
–12
A
Measuring mode
pH
Measuring
range
–13 - +20
Resolution
0.001
Measuring
accuracy
±0.003
(±1 digit, without sensor error, under reference conditions)
Measuring mode
U
Measuring
range
–1,200 - +1,200 mV
Resolution
0.1 mV
Measuring
accuracy
±0.2 mV
(±1 digit, without sensor error, under reference conditions)
11.1.2
Polarizer
A measuring input (Pol.) for polarizable electrodes.
Measuring mode
Ipol
Determination with adjustable polarization current.
Polarization
current
–120 - +120 µA (increment: 1 µA)
–125 - –121 µA / +121 - +125 µA: non-guaranteed values, dependent
on reference voltage +2.5 V
Measuring
range
–1,200 - +1,200 mV
Resolution
0.1 mV
Measuring
accuracy
±0.2 mV
(±1 digit, without sensor error, under reference conditions)