7 Parameters

848 Titrino plus

■■■■■■■■

103

Measuring mode Upol:
Input range

–125.00 - 125.00 µA

Selection

off

Default value

off

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

slow | optimal | fast | user

Default value

optimal

slow
For steep titration curves for which dosing must be carried out in small
steps at the endpoint.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For flat titration curves for which the endpoint is reached only slowly.
user
The individual titration parameters can be modified.

The settings of the individual titration rates are listed in table 3, page
104
.

Dynamics

This parameter is visible only when Titration rate = user.

This parameter defines the control range before the specified endpoint.
Individual volume steps are dosed in the control range, the dosing is finely
controlled. The closer the endpoint, the slower the dosing until the dosing
rate defined under Min. rate has been reached. The larger the control
range, the slower the titration. Outside of the control range, dosing is car-
ried out continuously, and the dosing rate is defined under Max. rate.

Measuring mode pH:
Input range

0.001 - 20.000

Default value

2.000

Selection

off

Measuring mode U and Ipol:
Input range

0.1 - 1250.0 mV

Default value

100.0 mV

Selection

off