beautypg.com

Metrohm 848 Titrino plus User Manual

Page 65

background image

■■■■■■■■■■■■■■■■■■■■■■

6 System settings

848 Titrino plus

■■■■■■■■

55

Selection

on | off

Default value

on

Temperature sensor

The instrument supports the use of two different temperature measure-
ment techniques:

NTC (Negative Temperature Coefficient)

Pt1000 (platinum resistance)

Select the type here that has been connected to the instrument. If you use
an NTC sensor, it is required that two characteristics for the sensor be
entered in addition. These characteristics are listed in the specifications of
the sensor.

Selection

Pt1000 | NTC

Default value

Pt1000

R (25 °C)

This parameter is visible only when Temperature sensor = NTC.

Nominal resistance of the NTC sensor at 25 °C.

Input range

1000 - 99999 ohm

Default value

30000 ohm

B value

This parameter is visible only when Temperature sensor = NTC.

Material constant of the NTC sensor. B values of NTC sensors are fre-
quently based on different reference temperatures (usually 25 °C and
50…100 °C).

Input range

1000 - 9999 K

Default value

4100 K