3 stop conditions, Stop conditions – Metrohm 848 Titrino plus User Manual
Page 84
7.1 Dynamic equivalence point titrations (DET)
■■■■■■■■■■■■■■■■■■■■■■
74
■■■■■■■■
848 Titrino plus
Input range
–125 - 125 µA (Increment: 1)
Default value
5 µA
U(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement. This parameter is available only
with U(pol) determinations.
Input range
–1250 - 1250 mV (Increment: 10)
Default value
400 mV
Electrode test
For polarizable electrodes, an electrode test can be carried out. A check is
made that the electrode is properly connected and that no short-circuit is
present. The electrode test is carried out when the determination is star-
ted. This parameter is available only with I(pol) and U(pol) determinations.
Selection
on | off
Default value
off
Stirrer
When this parameter is switched on, then the stirrer is switched on at the
start of the determination.
Selection
on | off
Default value
on
Stirring rate
Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 U/min. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 118. The optimum stirring
rate can be tested in the manual control.
The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:
■
"+": counterclockwise rotation
■
"–": clockwise rotation
Input range
–15 - 15
Default value
8
7.1.3
Stop conditions
Menu
▶ Parameters ▶ Stop conditions
The conditions for stopping a titration are defined under Stop condi-
tions.