7.2 Monotonic equivalence point titrations (MET)

■■■■■■■■■■■■■■■■■■■■■■

96

■■■■■■■■

848 Titrino plus

Curve

Curve report. The width of the curve is defined in the system settings (see
"Graphics width", page 64)
.

Selection

on | off

Default value

off

Calculations/Statistics

Output of the calculation formulas for the individual results. The results
are printed with the maximum resolution. This makes recalculation with
an external program possible. The following data is printed out in addition
when statistics is switched on:

Result and sample size of the individual determinations

Mean value as well as absolute and relative standard deviation

Selection

on | off

Default value

off

Measuring point list

Output of the measuring point list.

Selection

on | off

Default value

off

Parameters

All of the parameters of the current method are printed out with the para-
meter report.

Selection

on | off

Default value

off

PC/LIMS

The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 63)
.

The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt
.

Selection

on | off

Default value

off