2 start conditions, Start conditions – Metrohm 848 Titrino plus User Manual
Page 108
7.3 Endpoint titrations (SET)
■■■■■■■■■■■■■■■■■■■■■■
98
■■■■■■■■
848 Titrino plus
Input range
0.0 - 99.9 µL/min
Default value
0.0 µL/min
Cond. stop volume
Maximum permissible volume that can be dosed during conditioning.
Conditioning is stopped when the specified volume is dosed. If condition-
ing is continued by pressing [START] once again, then the titrant volume
that has already been dosed will not be taken into account; i.e. the dosing
starts again at zero. The stop volume should be adjusted to the size of the
titration cell in order to prevent any overflow.
Input range
0.00000 - 9999.99 mL
Default value
20.0000 mL
Selection
off
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Input range
0 - 999999 s
Selection
off
Default value
off
7.3.2
Start conditions
Menu
▶ Parameters ▶ Start conditions
The parameters that are carried out before the start of titration are
defined under Start conditions.
Activation pulse
Output of an activation pulse on a remote line. This activation pulse starts
a connected Dosimat.
Selection
on | off
Default value
off
Start delay time
Waiting time after the start of the determination, before titration takes
place. An auxiliary solution can, e.g. be added with a Dosimat during this
time (parameterization at the Dosimat). The Activation pulse parameter
must however be switched on for this purpose.
Input range
0 - 999999 s
Default value
0 s
Start volume
Volume that is dosed prior to the start of the titration.