13 manual control – Metrohm 848 Titrino plus User Manual
Page 54
5.13 Manual Control
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
848 Titrino plus
Results
Result report with determination properties,
sample data, calculated results, etc.
Curve
Curve report. The width of the curve is defined in
the system settings (see "Graphics width", page
64).
Measuring point
list
Measuring point list report.
Parameters
Report with all method parameters of the loaded
method.
System
System report with system settings, solution list,
external devices, etc.
Calculations/Statis-
tics
Calculation report. The statistics are also printed
out in the case of multiple determinations. The
individual determinations with the respective
sample size, the mean value, the absolute and
the relative standard deviation are printed out
for each result.
PC/LIMS
Machine-readable report with all of the data for
a determination. This report can be saved as a
TXT file to a connected USB flash drive or sent to
a terminal program or a LIMS via an RS-232
interface. The definition is made in the system
settings (see "PC/LIMS report", page 63).
Report as in
method
The reports that are defined in the method will
be printed out.
5.13
Manual Control
Menu
▶ Manual control
The following functions are available in the manual control:
■
Dosing
■
Measuring
■
Stirring
The available subfunctions are listed for each function in the function bar.