7.2 Monotonic equivalence point titrations (MET)
■■■■■■■■■■■■■■■■■■■■■■
84
■■■■■■■■
848 Titrino plus
Selection
off | ID1 | ID2 | ID1&ID2
Default value
off
Request sample size
If this parameter is switched on, then the value for the sample size will be
queried.
Selection
on | off
Default value
off
Request sample unit
If this parameter is switched on, then the unit for the sample size will be
queried.
Selection
on | off
Default value
off
Hold at request
If this parameter is switched on, then the method run will be paused dur-
ing the query. If the parameter is switched off, the titration will be started
in the background.
Selection
on | off
Default value
on
7.2.2
Titration parameters
Menu
▶ Parameters ▶ Titration parameters
The parameters that influence the sequence of the entire titration are
defined under Titration parameters.
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection
slow | optimal | fast | user
Default value
optimal
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted equivalence points.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For fast and less critical titrations.
user
The individual titration parameters can be modified.