beautypg.com

2 titration parameters, Titration parameters – Metrohm 848 Titrino plus User Manual

Page 94

background image

7.2 Monotonic equivalence point titrations (MET)

■■■■■■■■■■■■■■■■■■■■■■

84

■■■■■■■■

848 Titrino plus

Selection

off | ID1 | ID2 | ID1&ID2

Default value

off

Request sample size

If this parameter is switched on, then the value for the sample size will be
queried.

Selection

on | off

Default value

off

Request sample unit

If this parameter is switched on, then the unit for the sample size will be
queried.

Selection

on | off

Default value

off

Hold at request

If this parameter is switched on, then the method run will be paused dur-
ing the query. If the parameter is switched off, the titration will be started
in the background.

Selection

on | off

Default value

on

7.2.2

Titration parameters

Menu

Parameters Titration parameters

The parameters that influence the sequence of the entire titration are
defined under Titration parameters.

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

slow | optimal | fast | user

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted equivalence points.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For fast and less critical titrations.
user
The individual titration parameters can be modified.