Metrohm 848 Titrino plus User Manual
Page 82
7.1 Dynamic equivalence point titrations (DET)
■■■■■■■■■■■■■■■■■■■■■■
72
■■■■■■■■
848 Titrino plus
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed. This can be useful when the titration reaction pro-
ceeds slowly or the electrode is slow to respond.
Measuring mode Upol:
Input range
0.01 - 99.90 µA/min
Default value
50.00 µA/min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed. This can be useful when the titration reaction pro-
ceeds slowly or the electrode is slow to respond.
Min. waiting time
This parameter is visible only when Titration rate = user.
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
This parameter is visible only when Titration rate = user.
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Input range
0 - 999999 s
Default value
26 s
Table 1
Standard values for predefined titration rates for DET
Titration rate
slow
optimal
fast
Meas. point density
2
4
6
Min. increment
10.00 µL
10.00 µL
30.00 µL
Max. increment
off
off
off
Dosing rate
maximum
maximum
maximum