beautypg.com

Rockwell Automation 1734-IT2I Thermocouple and RTD Input Module User Manual User Manual

Page 61

background image

Publication 1734-UM004F-EN-E - December 2012

Calibrate Your Module 51

10. From the General Parameters dialog, click Accept Low Calibration.

11. From the General Parameters dialog, click the Download to Device

button.
The status indicator(s) for the channel(s) being calibrated blinks.

12. Set the resistance box to 500

Ω

. .

For the 1734-IR2E, set the resistance box to 200

Ω .

13. From the General Parameters dialog, click Accept High Calibration.