background image

51

Installation Standard Manual


Install the unit according the procedure described below (check options and special specifications
separately).

Model Serial

number Date

Person in charge of installation

(company name)

Person in charge of

installation

Judgment

Item

Method

Reference operation manual

Judgment

Specifications

1 Accessories

Check the quantities of accessories with
the quantities shown in the Accessory
column.

Specification

P.45

Visually check the surrounding area.

Caution: Pay attention to the ambient
environment.

Keep space.

Before Using This Unit

"2. Choose a proper place for
installation"

P.4

2 Installation

Pour water into the water bath.

Caution: Air release.

Before Using This Unit

"Installation Procedure"

P.7

Operation

Before Using This Unit
"1. Always ground this unit"

P.4

Before Using This Unit
"6. Choose a correct power
distribution board or receptacle"

P.5

1 Power

voltage

Using a tester, measure the voltage of
the voltage used by the customer
(distribution board, outlet, etc.).

Measure the voltage during operation
(the voltage must be within the
standard).

Caution: When a unit is to be connected
to the plug or breaker, use one that
conforms to the standard.

Specification

P.45

Before Using This Unit

"Installation Procedure"

P.7

2

Start of
operation

Start operation.
The circulating water must circulate.
Set the temperature at 20°C to confirm
the state.

Check: Water leakage is not permissible.

Operation Method

P.13

Description

1

Description of
operation

Explain the operation of each unit to the
customer according to this Operation
Manual.

All

2 Error

code

Explain error codes and the procedure for
resetting them to the customer according
to this Operation Manual.

In the Event of Failure…

P.42

3

Maintenance
inspection

Explain the operation of each unit to the
customer according to this Operation
Manual.

Maintenance Method

P.39

4

Completion of
installation
Information to
be entered

Enter the date of installation and the
name of the person in charge of
installation on the face plate on the
unit.

Enter necessary information on the

guarantee, and pass it to the customer.

Explain the after-sale service route to
the customer.

After Service and Warranty

P. 44

This manual is related to the following products: