Operation method, Calibration offset function – Yamato Scientific CLS600 Cool Line User Manual

Page 29

background image

26

Operation Method

Calibration Offset Function


Calibration offset is a function which corrects the difference between the temperature in bath and that of
controller (sensor temperature) if arises. The function parallel corrects the difference either to the plus or
minus side within the whole temperature range of unit. The function can be set or cancelled by the SUB
MENU key. "0" is set at factory shipment.

Start operation with the target setting temperature. Check the

temperature in bath with a thermograph after it is stabilized.

Check the difference between the setting temperature and that in

bath.

Press SUB MENU key for 4 seconds. Press SUB MENU key again.

Select "cAL" which means calibration offset, and press RUN/STOP
key.

Input the difference between setting temperature and inner bath

temperature by ▼ ▲ key, and press SUB MENU key for a few
seconds to complete the setting.

™

When the offset correction temperature is set to the minus side, the

temperature on the measurement temperature display screen falls by
the setting temperature, while the temperature on bath rises.
When it is set to the minus side, the temperature on the measurement
temperature display screen rises by the setting temperature, while the
temperature on bath falls.

™

The unit has two-point correction function, which performs offset

between low-temperature zone and high-temperature zone.

™

Please consult our local branch office when carrying out validation of

temperature controller.

Current temperature

Corrected temp. to minus side

Corrected temp. to plus side