Operation method, Quick auto stop operation – Yamato Scientific CLS600 Cool Line User Manual
Page 21

18
Operation Method
Quick Auto Stop Operation
Quick auto stop
operation procedure
This operation is used to specify the period up to automatic stop, i.e.,
sets the auto stop timer during operation.
1. Enter the quick auto stop mode during fixed temperature
operation
• Confirm that the FIXED TEMP. lamp is lit, and it is under
operation.
• Press
TIMER
key.
• AUTO STOP lamp will blink.
2. Set the timer
• Setting value will blink at setting temperature screen.
• Set the proper time by ▼▲ key.
The setting value will be smaller by ▼ key, and larger by ▲
key.
Timer function:
• The maximum setting time is "999 hours and 50 minutes".
• The time can be set in increments of a minute under 99 hours
and 59 minutes.
• It can be set in increment of ten minutes over 100 hours.
• The
"▼▲" can change the setting time quickly when it is pressed
continuously. Press them discontinuously when fine adjustment
is needed.
3. Start timer operation
• Press
RUN/STOP
key.
• Start timer operation with the FIXED TEMP. lamp and AUTO
STOP lamp on.
• Timer operation starts when RUN/STOP key is pressed.
• Remaining time is displayed at the setting temperature screen