beautypg.com

Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual

Page 54

background image

54

Advanced functions

3. Programme the chamber on the computer. The chosen settings are written and stored on

the card.

Via the “Celsius“ software, the MEMoryCard XL can be provided with write protection.

The programme on the card can then no longer be altered on the controller.
Programme alterations via the controller are only possible if the write protection has

been disabled from the computer with the “Celsius“ software.

Programming the MEMoryCard XL from the computer with the read-write device

Using the read-write device which can be purchased separately ( Fig. 33 ), the MEMoryCard XL

can be programmed offline on the computer with “Celsius“ without the need for a constant

climate chamber being connected. When you do this, make sure that the contacts on the

MEMoryCard XL are pointing upwards to the markings on the read-write device.

Fig. 33

Read-write device

The programme remains stored on the MEMory-Card XL even after it is removed from

the appliance. It can also be overwritten at any time with “Celsius“ via the computer.

Logging on the MEMoryCard XL

During a programme sequence running from the chip card, the humidity and temperature

setpoints are continually logged on the memory card. These can be read in via “Celsius” after

the programme is finished and printed out. This is described in the “Celsius“ user manual.
There is a certain amount of storage space available for logging, depending on the pro-

gramme duration. The scanning rate is set automatically by the controller, depending on the

programme duration.
The actual value can be logged on the MEMoryCard XL at one minute intervals

for a programme lasting up to 135 hours. For programmes lasting longer, the scanning rate is

lengthened to a maximum of 30 min.

At each programme start, the logging is restarted and old log data is overwritten.