beautypg.com

4 basic equipment – Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual

Page 11

background image

11

Design and description

Range A:

In this range, temperature and humidity can be combined as you please, without resulting in

any significant condensation. If there are extreme ambient conditions, the working range may

be restricted.

Range B:

If the specified range is exceeded upwards, e.g. 80 % rh at 60°C, the hot steam fed in will im-

mediately condense, due to the dew-point, at the coldest point in the appliance.

Range C:

At low temperatures and low relative air humidity, the effective range is heavily dependent on

the degree of humidity of the chamber load.

2.4

Basic equipment

Low-noise, long-life and energy-saving Peltier cooling and heating technology (in heat-

ing mode, part of the energy required is extracted from the surroundings = heat pump

principle)

Electronic fuzzy-supported PID process controller with permanent performance adjust-

ments and time-saving self-diagnosis system to quickly locate errors (see pages 38 and 45)

Alphanumeric text display

Control of the appliance and logging the actual values with a preformatted empty

MEMoryCard XL (storage capacity 32 kByte, reprogrammable for up to 40 ramp segments

and in addition 135 hours of log memory for temperature and humidity with a scanning

interval of 1 minute)

Programme sequence control for up to 40 ramp segments

Integrated week time switch with group function (e.g. all working days)

Retracting push-turn control for simple operation

Optical alarm display

Integrated horn as alarm if limit values are crossed, as acoustic signal when programme

ends and to confirm input (button click)

Digital monitoring control for overtemperature, undertemperature and automatic tem-

perature monitor (ASF)

Two separate Pt100 temperature sensors DIN class A in a 4-wire circuit for control and

monitoring

Parallel printer port (PCL3-compatible)

USB interface for computer-based temperature control programmes and to read out the

controller’s internal log memory

MEMMERT software "Celsius" for remote control of the chamber via computer (see page

37) and to read out the internal controller log memory

As an optional accessory, external card reader for MEMory-Card XL for connection to the

PC-USB interface, printer cable (parallel, shielded)- 25-pin

Language setting (German, English, French, Spanish, Italian)

Capacitive humidity sensor

Active humidity control guarantees that setpoint humidity is quickly achieved, with short

recovery times, and avoiding condensation formation