Operation and control

9. Set the close statement of the first ramp segment:

Turn the push/turn control to the right until a segment

close statement, e.g. end, appears.

Hold down the SET key and set the close statement SPWT

[TH] with the push/turn control.

10. Set the duration of the second ramp segment:

Turn the push/turn control to the right until the time

display flashes.

Hold down the SET key and adjust the time to 0:45 with

the push/turn control.

h

11. Set the temperature of the second ramp segment:

Turn the push/turn control to the right until the tem-

perature display flashes.

Hold down the SET key and set the desired temperature

setpoint to 37.0 °C with the push/turn control.

°C

12. Setting the light intensity of the second ramp segment

(optional):

Turn the push-turn control to the right until the lighting

display flashes.

Hold down the SET key and set the desired light intensity

of 80 % (eight bars are lit up).

13. Set the relative humidity of the second ramp segment:

Turn the push/turn control to the right until the humidity

display flashes.

Hold down the SET key and set the desired humidity

setpoint to 70.0 % rh with the push/turn control.

%

rh

14. Set the close statement of the second ramp segment:

Turn the push/turn control to the right until a segment

close statement, e.g. end, appears.

Hold down the SET key and set the close statement next

with the push/turn control.

15. Set the time for the third ramp segment:

Select the time display with the push/turn control.

Hold down the SET key and set the time 1:00 with the

push/turn control.

h

16. Set the temperature of the third ramp segment:

Turn the push/turn control to the right until the tem-

perature display flashes.

Hold down the SET key and set to 15.0 °C with the push/

turn control.

°C