6 operating mode settings – Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual
Page 25

25
Operation and control
5.6
Operating mode settings
1. Switch on the appliance by pressing the main switch (main switch comes out of appliance,
see page 23 ).
2. Keep the SET key depressed for approx. three seconds and the selected mode
starts flashing.
3. Select the desired operating mode/function (normal mode, week time switch,
programming mode, printer or basic appliance settings) by turning control
with SET key held down.
4. Release the SET key, and the selected operating mode is saved.
5.6.1 Normal mode
In this operating mode, the constant climate chamber runs in permanent operation at the
values set on the operating panel (see above). Example of settings: See next chapter 5.6.2 .
1. Load the constant climate chamber (see page 22 ).
2. Switch on appliance. To do this, press the push-turn control on the operating panel so that
it comes out of the appliance (see page 23 ).
3. Select the normal operating mode
with the push/turn control:
SETUP
4. As described above, set the individual parameters with the push-turn control and the SET
key:
Temperature setpoint
Adjustment range: 0°C to 70°C
°C
Interior light (optional)
Adjustment range: 0 % to 100 % in steps of
10%
The interior lighting can only be acti-
vated at a working temperature of up
to 40 ºC.
t2
Temperature monitoring
Adjustment range:
MIN MAX AUTO
(see also page 43 )
MIN
MAX
AUTO
°C
Humidity setpoint
Adjustment range: 10 to 90 %rh,
off
rh
%
Not all combinations of temperature and humidity are possible (see also page 10 ).
set
set
set