beautypg.com

Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual

Page 34

background image

34

Operation and control

5.6.7 Settings example programming mode

On Monday at 8:00, the constant climate chamber should heat up to 37 °C, with 50 % light

intensity (optional), as quickly as possible and reach a relative humidity of 70 % rh. Once the

temperature and humidity have been reached, the constant climate chamber should retain

the setpoint values for 45 minutes at 80 % light intensity and then cool down within one

hour, at a light intensity of 30 %, to a humidity of 50 % rh and 15 °C ( Fig. 21 ).

Mo 8.00 h

60 °C

40 °C

20 °C

50 %

30 %

80 %

0.45 h

1.00 h

t

0.01 h

70 % rh

50 % rh

Ramp 1

Ramp 2

Ramp 3

Fig. 21

Settings example programming mode

Before programming ramp sequences, especially before programming complicated ones,

it is recommended that you prepare a similar plan to ensure that you enter the required

ramp commands correctly, as described below. For better orientation, it is recommended

that you programme large programmes graphically on the computer using the "Celsius"

software.