48

Advanced functions

7.5 Calibration

7.5.1 Temperature calibration

The constante climate chamber can be calibrated customer-specifically using three calibration

temperatures of your choice:

CAL.1

temperature calibration at low temperature

CAL.2

temperature calibration at medium temperature

CAL.3

temperature calibration at high temperature

For each selected balance point ( Fig. 27 ) a positive or negative compensation correction value

can be set between –4.9 ºC and +4.9 ºC. There must be a difference of at least 10 K between

the individual balance points.

CAL 1

+0,5°C

CAL 3

+0,8°C

CAL 2

-0,4°C

factory

calibration

37°C

20°C

0°C

5°C

Fig. 27

Temperature calibration (example)

Setting:
1. Set the desired compensation temperature in the SETUP (see page 42 ) and set the accom-

panying compensation correction value to 0.0 °C.

2. With a reference instrument, measure the deviation in the stationary state in the selected

compensation temperature.

3. Set the compensation correction value in the SETUP. If the measured reference tempera-

ture is too low, the compensation correction value must be set with a negative sign.

4. Perform a control measurement with the reference instrument.
5. Repeat the procedure for the other two balance points if necessary.