beautypg.com

Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual

Page 15

background image

15

Design and description

Model

HPP 108

HPP 746

Max. load per sliding shelf [kg]

30

30

Max. load per appliance [kg]

60

160

Electrical equipment

Operating voltage

see nameplate

Current consumption

see nameplate

Protection class

1 (acc. to EN 61010)

Protection type

IP 20 (EN 60529)

Interference-suppressed

Class B acc. to EN 55011

Fuses

Appliance: 250 V/15 A (appliance)

Controller: 100 mA (230 V) or 200 mA (115

V)

Temperature

Temperature recording: Pt100 in a 4-wire

circuit

Adjustment range: 0 °C to 70 °C

Adjustment precision: 0.1 °C

Working temperature range: 5 °C to 70 °C

Temperature variation (time): ±0.1 °C

Spatial deviation: max. ±0.3 °C bei 37°

Monitoring

Temperature recording: Pt100 in a 4-wire

circuit

Settings range -5 °C to +75 °C

Adjustment precision: 0.1 °C

Humidity

The relative humidity in the chamber is

measured by a capacitive humidity sensor and

displayed digitally in percent

Measurement precision of the humidity

sensor: 0.5 % rh

Adjustment range: 10 bis 90 % rh, off*

Adjustment precision: 1 % rh

Display range: 1 bis 98 % rh

Temperature variation (time): max. ±1.5

% rh

* Setting off: Humidity control disabled

* see Fig. 5 on page 16 .