Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual
Page 32

32
Operation and control
After releasing the SET key ...
►
... a new programme can be created
as described above, or an existing
one be edited
EDIT
►
... the programme can be stopped
STOP
►
... the programme can be started
START
5.6.6 Close statements for ramp segments
Each ramp must be completed with a close statement connecting the ramp to the next one.
These commands thus control the programme sequence:
NEXT
Connect the next programme segment.
SET-POINT WAIT (T – temperature)
Wait until setpoint has been reached.
Appliance starts the next programme segment only when the
programmed setpoint temperature has been reached, even if the
set heating up time has already elapsed.
SET-POINT WAIT (H – humidity)
Wait until setpoint humidity has been reached.
Appliance starts the next programme segment only when the
programmed setpoint humidity has been reached, even if the set
heating up time has already elapsed.
SET-POINT WAIT (TH – temperatur and humidity)
Wait until the setpoint temperature and setpoint humidity has
been reached.
Appliance starts the next programme segment only when the
programmed setpoint temperature and programmed setpoint
humidity have been reached, even if the set heating up time has
already elapsed.
Ramp repeat function
The programme entered is repeated after all programmed seg-
ments have been run.
1-99 = repeats
cont = endless loop repeat function
Programme end while maintaining the temperature and humidity
of last programme ramp
Programme end, switching off the heating /cooling function and
humidity