Operation and control
17. Setting the light intensity of the third ramp segment
(optional):
Turn the push-turn control to the right until the lighting
display flashes.
Hold down the SET key and set the desired light intensity
of 30 % with the push-turn control (three bars are lit
up).
18. Set the relative humidity of the third ramp segment:
Turn the push/turn control to the right until the humidity
display flashes.
DHold down the SET key and set the desired humidity
setpoint to 50.0 % rh with the push/turn control.
%
rh
19. Adjust the close statement of the third ramp segment:
Turn push/turn control to the right until a close state-
ment appears in the display, e.g. end, and press the SET
key briefly to confirm.
20. Leave the EDIT programme write mode:
Turn push/turn control to the right until EXIT appears in
the display, and press the SET key briefly to confirm.
21. Set temperature monitoring:
Turn the push-turn control to the right and adjust the
temperature monitoring (for more detailed information
on this, see page 43 ).
°C
MIN
AUTO
MAX
22. Start programme:
Turn the push/turn control to the right until the stop icon
flashes.
Hold down the SET key and select the start icon with
the push-turn control. Release the SET key, and the pro-
gramme is started.