Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual
Page 35

35
Operation and control
1. Setting the programme operating mode:
Hold the SET key down for approx. three seconds; the
current operating mode then begins to flash. Select the
programme operating mode with the push/turn control
while holding down the Set key.
After releasing the SET key, the controller is in the pro-
gramme mode.
SETUP
2. Edit programme:
Select EDIT with the push/turn control while holding
down the SET key.
After releasing the SET key, the controller is in the pro-
gramme write mode.
3. Weekday for delayed programme start:
Set the start day MO by turning the push-turn control
while the SET key is held down.
Mo
4. Set the time for delayed programme start:
Select the time display with the push/turn control.
Hold down the SET key and adjust the time to 8:00 with
the push/turn control.
h
5. Set the duration of the first ramp segment:
Turn the push/turn control to the right until the time
display flashes.
Hold down the SET key and adjust the time to 0:01 with
the push/turn control.
h
6. Set the temperature of the first ramp segment:
Turn the push/turn control to the right until the tempera-
ture display flashes.
Hold down the SET key and set the desired temperature
setpoint to 37.0 °C with the push/turn control.
°C
7. Setting the light intensity of the first ramp segment
(optional):
Turn the push-turn control to the right until the lighting
display flashes.
Hold down the SET key and set the desired light intensity
of 50 % (five bars are lit up).
8. Set the relative humidity of the first ramp segment:
Turn the push/turn control to the right until the humidity
display flashes.
Hold down the SET key and set the desired humidity
setpoint to 70.0 %rh with the push/turn control.
%
rh