Rotation circuit, Row marker operation, Rotation circuit -4 – Kinze 3660 Lift and Rotate Planter Rev. 7/14 User Manual
Page 144: Row marker operation -4

M0217-01
Model 3660
6-4
Rev. 7/11
TM
Troubleshooting
ROTATION CIRCUIT
PRObLEM
POSSIbLE CAUSE
SOLUTION
Cylinder does not extend, but
will retract.
Solenoid valve coil in port V12 defective.
Switch coil from port V12 with coil in port
V9. If cylinder extends but will not retract,
replace defective coil from port V12.
Solenoid valve cartridge in port V12 stuck
closed.
Switch cartridge from port V12 with
cartridge in port V9. If cylinder extends but
will not retract, replace defective cartridge
from port V12.
Cylinder does not retract, but
will extend.
Solenoid valve coil in port V9 defective.
Switch coil from port V9 with coil in port
V12. If cylinder extends but will not retract,
replace defective coil from port V9.
Solenoid valve cartridge in port V9 stuck
closed.
Switch cartridge from port V9 with
cartridge in port V12. If cylinder extends
but will not retract, replace defective
cartridge from port V9.
ROW MARKER OPERATION
PRObLEM
POSSIbLE CAUSE
SOLUTION
Right marker lowering slower
than left marker.
Solenoid valve cartridge in port V1 not
opening completely.
Switch with cartridge in port V2. If problem
repeats, replace cartridge.
Hose pinched or collapsed.
Inspect hose routing. Replace or repair
hoses as required.
Left marker lowering slower
than right marker.
Solenoid valve cartridge in port V2 not
opening completely.
Switch with cartridge in port V1. If problem
repeats, replace cartridge.
Hose pinched or collapsed.
Inspect hose routing. Replace or repair
hoses as required.
Both markers lowering.
Solenoid valve cartridge stuck open. If left
marker switch is selected, right cartridge
(V1) is defective. If right marker switch is
selected, left cartridge (V2) is defective.
Replace solenoid valve cartridge.
Neither marker lowers.
Blown fuse.
Check red light on control console. It
should be on if switch is on. If light is not
on, switch to opposite marker position. If
light comes on, switch may be defective.
Replace switch. Otherwise replace fuse.
Coils at V1 and V2 not energized.
Poor ground on wire, bad wire connection
or damaged wire. Repair as required.
Marker flow control valve closed too far.
See “Row Marker Speed Adjustment”.
Neither marker will raise.
Marker flow control valve closed too far.
See “Row Marker Speed Adjustment”.
Right marker will not lower.
Solenoid coil in port V1 not energized.
Check switch on control console. Replace
if defective. Check coil ground wire. Check
for poor connection or damaged wire.
Solenoid cartridge in port V1 stuck closed.
Switch cartridge with one on the planter
you know is operating properly. If right
marker lowers, replace defective cartridge.
Left marker will not lower.
Solenoid coil in port V2 not energized.
Check switch on control console. Replace
if defective. Check coil ground wire. Check
for poor connection or damaged wire.
Solenoid cartridge in port V2 stuck closed.
Switch cartridge with one on the planter
you know is operating properly. If left
marker lowers, replace defective cartridge.
Markers traveling too fast
and damaging rubber stop
on transport stands and/or
damaging pivot at rod end of
marker cylinders.
Marker transport stand not adjusted
correctly to allow marker cushion cylinders
to operate as designed.
See “Row Marker Transport Stand
Adjustment”.
Adjust row marker flow control valve.
See “Row Marker Speed Adjustment”.