3 indicators and switches, 1 led, 2 led – INFICON BPG402-SE ATM to Ultra-High Vacuum Gauge (EtherCAT) User Manual

Page 9: 3 led ( port), 4 led ( port), 5 device address switch

background image

tira93e1 (2014-04) BPG402-SE.cp

9



Displays the operating status.

Color

LED State

Description

off

INIT (initialization status) or no power applied to device.

blinking

(200 ms on
200 ms off)

PREOP (pre-operational status).

single flash

(200 ms on

1000 ms off)

SAFEOP (safe-operational status).
Communication of cyclic data transfer running. Input values
available, output values written to the device but not up-
dated on device output.

green

on

OP (operational status).

Displays the error content.

Color

LED State

Description

off

No error or no power applied to device.

blinking

(200 ms on
200 ms off)

Error occurred (see error parameter).

single flash

(200 ms on

1000 ms off)

Slave device application has changed the EtherCAT state
autonomously, due to local error (see error parameter).

double flash

(200 ms on

200 ms off
200 ms on

1000 ms off)

An application watchdog timeout has occurred.
Sync Manager Watchdog timeout or communication
timeout occurred.

red

on

A critical communication or application controller error has
occurred. Application controller is not responding any more
(PDI Watchdog Timeout detected by ESC)

Displays the input status.

Color

LED State Description

off

Port not connected or no power applied to device.

blinking

Port connected and communication active.

green

on

Port not connected but no communication.

Displays the output status.

Color

LED State Description

off

Port not connected or no power applied to device.

blinking

Port connected and communication active.

green

on

Port not connected but no communication.

During device initialization, the device address switches are read by the device
firmware. This device address is supported to the master as Explicit Device
Identification.

Example: Value of the Explicit Device ID = 0xDD (dec 221):
0x10 * 0xD (dec 208) + 0x1 * 0xD (dec 13)

3.3 Indicators and Switches

3.3.1 LED

RUN

3.3.2 LED

ERR

3.3.3 LED ( Port)

LA

IN

3.3.4 LED ( Port)

LA

OUT

3.3.5 Device Address Switch