19 exception reset command, 20 store parameters command – INFICON BPG402-SE ATM to Ultra-High Vacuum Gauge (EtherCAT) User Manual

Page 30

background image

30

tira93e1 (2014-04) BPG402-SE.cp

Status (supported values)

0
1
2
3

255

Reserved
Reserved
Last command completed, error, no response
Reserved
Command is executing

Response
Byte 0

See Subindex 0x02

Byte 1

Unused

Execution of this command clears the latched exceptions.

Index

SI

DataType

NV Access PM Name

FBF1 0x01

V_STRING(5) RW Command

0x02

BYTE

RO

Status

0x03

V_STRING(2)

RO

Response

A exception reset is initiated when the following byte sequence is sent.

Command
Byte 0
Byte 1
Byte 2
Byte 3
Byte 4

0x74
0x65
0x73
0x65
0x72

Status (supported values)

0
1
2
3

255

Last command completed, no error, no response
Reserved
Last command completed, error, no response
Reserved
Command is executing

Response
Byte 0

See Subindex 0x02

Byte 1

Unused


Execution of this command will store all parameters to non-volatile memory. If a
device automatically saves all non-volatile parameters at the time they are set, this
command will not take any action.

Index

SI

DataType

NV Access PM Name

FBF2 0x01

V_STRING(4) RW Command

0x02

BYTE

RO

Status

0x03

V_STRING(2)

RO

Response

Subindex 0x02

Subindex 0x03

4.7.19 Exception Reset

Command

Subindex 0x01

Subindex 0x02

Subindex 0x03

4.7.20 Store Parameters

Command