Indicates whether the Value parameter contains a value in over range.
Reading Valid Module 1
0
No Overrange Exceeded
1 Overrange
Exceeded
Indicates whether the Value parameter contains a value in under range.
Reading Valid Module 1
0
No Underrange Exceeded
1 Underrange
Exceeded
Index
SI
DataType
NV Access PM Name
6010
0x0E BOOL
RO
tx
TxPdoState Module 2
0x11 REAL
RO
tx
Sensor Value Module 2
Is set if the device is not in Safe State (value (I 0x6010, SI 0x11) = valid)
TxPdoState Module 2
0
1
Invalid
Valid
The corrected, converted, calibrated final analog input value of the sensor.
Index
SI
DataType
NV Access PM Name
6015
0x01 BOOL
RO
tx
Reading Valid Module 2
0x02
BOOL
RO
tx
Overrange
Exceeded
Module 2
0x03
BOOL
RO
tx
Underrange
Exceeded
Module 2
0x05 BOOL
RO
tx
Emission Status Off/On
Module 4
0x06 BOOL
RO
tx
Degas Status Off/On
Module 4
Indicates whether the Value parameter contains a valid value within the specified
accuracy or not.
Reading Valid Module 2
0 Invalid
1 Valid
Indicates whether the Value parameter contains a value in over range.
Reading Valid Module 2
0
No Overrange Exceeded
1 Overrange
Exceeded
Subindex 0x02
Subindex 0x03
4.4.3 Input Common
Subindex 0x0E
Subindex 0x11
4.4.4 Input Hot Cathod Ion
Subindex 0x01
Subindex 0x02