1 technical data, Ethercat interface dimensions [mm – INFICON BPG402-SE ATM to Ultra-High Vacuum Gauge (EtherCAT) User Manual
Page 6
![background image](/manuals/562581/6/background.png)
6
tira93e1 (2014-04) BPG402-SE.cp
1 Technical
Data
Further technical data
→ [1], [2], [3].
Communication protocol
protocol specialized for EtherCAT
Communication standards
ETG.5003 Part 1: Common Device Profile (CDP)
ETG.5003 Part 2080: Specific Device Profile
(SDP) Vacuum Pressure Gauge
Data rate
100 Mbps
Node address
Explicit Device Identification
Physical layer
100BASE-Tx (IEEE 802.3)
Digital functions
read pressure, select units (Torr, mbar, Pa)
emission control, degas function
monitor gauge status, filament status
safe state allows definition of behavior in case of
error
detailed alarm and warning information
Analog functions
0 … 10 V analog output pressure indication
two setpoint relays A + B
Setpoint relays
Range
Relay contact
Hysteresis
Contact rating
2
1×10
-9
… 100 mbar
n.o. potential free
10 % of reading
≤30 V / ≤0.5 A (dc)
EtherCAT connector
2 × RJ45, 8-pin (socket)
input
Cable
shielded, special Ethernet Patch Cable or
Crossover Cable (CAT5e quality or higher)
Cable length
≤100 m
Process data
Fixed PDO mapping and configurable PDO
mapping
Mailbox (CoE)
SDO requests, responses and information
DN 40 CF-R
48
DN 25 ISO-KF
42
111
59
LA
LA
PWR
ERR
RUN
OUT
IN
FIL
67
EtherCAT interface
Dimensions [mm]