7 latched device warning details, 8 latched device error details – INFICON BPG402-SE ATM to Ultra-High Vacuum Gauge (EtherCAT) User Manual
Page 24

24
tira93e1 (2014-04) BPG402-SE.cp
Active Device Error Details Module 2 (Index F383)
Bit 0
0
Bit 1
Electronics failure
Bit 2…31
0
Active Manufacturer Error Details (Index F384)
Bit 0
Internal Communication Exception (between
EtherCAT-Protocol and Device-Application)
Bit 1
Wrong software version used in base device
Bit 2
Wrong sensor type detected
Bit 3…31
0
Index
SI
DataType
NV Access PM Name
F390 USINT RO tx
Latched Exceptions Status
F391
0x01 UDINT
RO
tx
Latched Device Warning
Details Device
0x02 UDINT
RO
tx
Latched Device Warning
Details Module 1
0x03 UDINT
RO
tx
Latched Device Warning
Details Module 2
F392 0x01
UDINT RO tx Latched
Manufacturer
Warning Details
Latched version of 0xF380.
Latched Device Warning Details Device: Latched version of 0xF381:01.
Latched Device Warning Details Module 1: Latched version of 0xF381:02.
Latched Device Warning Details Module 2: Latched version of 0xF381:03.
Latched Manufacturer Warning Details: Latched version of 0xF382:01.
Index
SI
DataType
NV Access PM Name
F393
0x01 UDINT
RO
tx
Latched Device Error
Details Device
0x02 UDINT
RO
tx
Latched Device Error
Details Module 1
0x03 UDINT
RO
tx
Latched Device Error
Details Module 2
F394 0x01
UDINT RO tx Latched
Manufacturer
Error Details
Latched Device Error Details Device: Latched version of 0xF383:01.
Latched Device Error Details Module 1: Latched version of 0xF383:02.
Latched Device Error Details Module 2: Latched version of 0xF383:03.
Latched Manufacturer Error Details: Latched version of 0xF384:01.
Subindex 0x03
Subindex 0x01
4.7.7 Latched Device Warning
Details
Latched Exceptions Status
Subindex 0x01
Subindex 0x02
Subindex 0x03
Subindex 0x01 (F392)
4.7.8 Latched Device Error
Details
Subindex 0x01
Subindex 0x02
Subindex 0x03
Subindex 0x01 (F394)