tira93e1 (2014-04) BPG402-SE.cp

Active Device Error Details Module 2 (Index F383)
Bit 0

0

Bit 1

Electronics failure

Bit 2…31

0

Active Manufacturer Error Details (Index F384)
Bit 0

Internal Communication Exception (between
EtherCAT-Protocol and Device-Application)

Bit 1

Wrong software version used in base device

Bit 2

Wrong sensor type detected

Bit 3…31

0


Index

SI

DataType

NV Access PM Name

F390 USINT RO tx

Latched Exceptions Status

F391

0x01 UDINT

RO

tx

Latched Device Warning
Details Device

0x02 UDINT

RO

tx

Latched Device Warning
Details Module 1

0x03 UDINT

RO

tx

Latched Device Warning
Details Module 2

F392 0x01

UDINT RO tx Latched

Manufacturer

Warning Details

Latched version of 0xF380.

Latched Device Warning Details Device: Latched version of 0xF381:01.

Latched Device Warning Details Module 1: Latched version of 0xF381:02.

Latched Device Warning Details Module 2: Latched version of 0xF381:03.

Latched Manufacturer Warning Details: Latched version of 0xF382:01.

Index

SI

DataType

NV Access PM Name

F393

0x01 UDINT

RO

tx

Latched Device Error
Details Device

0x02 UDINT

RO

tx

Latched Device Error
Details Module 1

0x03 UDINT

RO

tx

Latched Device Error
Details Module 2

F394 0x01

UDINT RO tx Latched

Manufacturer

Error Details

Latched Device Error Details Device: Latched version of 0xF383:01.

Latched Device Error Details Module 1: Latched version of 0xF383:02.

Latched Device Error Details Module 2: Latched version of 0xF383:03.

Latched Manufacturer Error Details: Latched version of 0xF384:01.

Subindex 0x03

Subindex 0x01

4.7.7 Latched Device Warning

Details

Latched Exceptions Status

Subindex 0x01

Subindex 0x02

Subindex 0x03

Subindex 0x01 (F392)

4.7.8 Latched Device Error

Details

Subindex 0x01

Subindex 0x02

Subindex 0x03

Subindex 0x01 (F394)