Compound elution order – INFICON Explorer Portable Gas Chromatograph User Manual
Page 293
Explorer Operator Manual
291
Compound Elution Order
The following tables list the elution order and typical retention times for all Assay 1
compounds on each column. This information will be needed if you are customizing
the standard method. Heavy compounds, such as 1,2-Dichlorobenzene, are not
generally analyzed on the more retentive Columns B and C, since retention times
will be unacceptably long.
The information in these tables is for reference only. Retention times will vary on
your instrument, and exact retention times for your instrument, are contained in the
Assay1.app file.
Compounds with retention times differing by 6% or less are defined as co-elutions
and are indicated as such in the tables. Some compounds identified as co-elutions
may be resolved on your instrument.
Also indicated are the compounds that are optimally detected with an ECD, these
compounds may also be detected on the PID with reduced sensitivity.
Column A Elution Order
Compound
ECD
R.T. (S)
Co-elution
Vinyl Chloride
59
Yes
Chloromethane
59
Yes
Bromomethane
63
Yes
Acetone
68
Yes
1,1-Dichloroethylene
68
Yes
Dichloromethane
72
Yes
Carbon Disulfide
73
Yes
Chloroethane
73
Yes
t-1,2-Dichloroethylene
74
Yes
Vinyl Acetate
77
Yes