INFICON Explorer Portable Gas Chromatograph User Manual
Page 105
Explorer Operator Manual
103
Follow the instructions below:
1. Make a copy of the assay disk, and store the original.
2. Create a new directory for your assay disk.
If you have more than one instrument, create a directory for each instrument.
Assay disks are specific to the instrument with the matching serial number.
3. Copy the contents of the assay disk into this directory.
Each assay you have ordered will be supplied on one 3.5” disk. Refer to the
assay notes for details of the files on the disk.
4. Connect
Explorer to a serial port on the computer.
5. Turn on
Explorer and allow it to warm up.
6. Start SiteChart.
7. Go to the Assay tab and open the assay.
Depending on which assay you are loading, you may be prompted, by Assay
Wizard, to configure your library with respect to co-eluting compounds. See
Assay Wizard on page
104.
8. Check the assay settings and then click on the Send to
Explorer button.
If assays created on one instrument are downloaded to a second instrument an
error message will be displayed. You must update the library for the second
instrument as outlined in Updating Library Information on page
110.
Once the file has been downloaded,
Explorer has been set up for your assay.
9. Calibrate
Explorer.
NOTE: You must calibrate
Explorer so that compound concentrations will be accurately
reported.