2. When the run is complete, review the chromatogram and peak report to ensure
that all compounds have been detected and that the reported concentrations are
correct.
3. Select the Assay tab and ensure all library concentrations are correct.
4. Open the Analyzer menu and select
Calibrate
.
SiteChart will calibrate
Explorer. The Explorer library will be updated
accordingly.
Explorer is now calibrated and ready for operation.
The information in the SiteChart library will not be updated. Each time you
load a new assay and download it to the instrument, you must calibrate the
instrument in order to update the current library information.
Service
The
Service
command displays a window that shows the
Explorer display. This
can be used to troubleshoot instrument problems remotely. This command allows
you to access all
Explorer functions from your computer.
NOTE:
Explorer cannot be powered up from this window. If Explorer is powered down, then
you must turn it on by pressing the On/Off button on the instrument.
The buttons control
Explorer in the same manner as the keys on Explorer.
Communication Menu
The Communications menu allows you to setup communications between the
computer and the instrument.
The communication pull-down menu has the following options:
Setup
allows you to configure which serial port will be used to connect the
instrument to the remote computer.
Dial Out
allows you to dial out to the specified number and establish
communication with a remote computer. This feature will allow you to dial out
to a remote instrument from SiteChart.